User Manual 1178.5566.02 ─ 09
Returns the error number, but no error description.
Return values:
<Error>
<n>
Number as defined in the SCPI standard
0
No errors have occurred since the error queue was last read out.
Usage:
Query only
SYSTem:ERRor:COUNt?
Queries the number of entries in the SCPI communication error queue.
Return values:
<ErrorCount>
Total number of entries
Example:
SYSTem:ERRor:COUNt
Queries the number of entries in the error queue.
Response: 1
One error has occurred since the error queue was last read out.
Usage:
Query only
SYSTem:ERRor:EXTended[:STATe]
<state>
Enables or disables the extended error messages that provide more information than
. The location of errors in the parsed command strings is
also shown.
Parameters:
<state>
*RST:
ON
SYSTem:ERRor[:NEXT]?
Queries the SCPI communication error queue for the oldest entry and removes it from
the queue.
Returns an error number and a short description of the error.
Return values:
<ErrorCode>
<n>
Number as defined in the SCPI standard
0
No errors have occurred since the error queue was last read out.