Remote control commands
R&S
®
NRX
396
User Manual 1178.5566.02 ─ 09
Example:
MEM:STAT:MAP? "test"
Manual operation:
See
See
MEMory:STATe:RESet
Resets the setup names to factory default, "Setup 1", "Setup 2" and so on.
Usage:
Event
MEMory:TABLe:DATA?
Queries the content of the selected table.
Use
The response consists of data pairs (frequency - offset):
<frequency 1>,<offset 1>,<frequency 2>,<offset 2>,<frequency 3>,<offset 3>, ...
Frequency in Hz, offset in dB.
Example:
MEM:TABL:DATA?
Query
1.02,0.00,2.03,
0.00,3.04,0.00
Response
Usage:
Query only
MEMory:TABLe:DATA:POINts?
Queries the number of data pairs (frequency - offset) in the selected table.
Use
Usage:
Query only
MEMory:TABLe:FREQuency
<value>...
Defines the frequency values in the selected correction table. Existing data is overwrit-
ten. Take care that the number of frequency values matches the number of offset val-
ues, defined by
. If the numbers differ, excess
values are ignored.
Use
Parameters:
<value>
Numeric values with a maximum of 2
digits after the decimal
point, separated by commas. Values with more than 2
decimal
places are rounded.
Default unit: Hz
Managing setups and correction tables