D) Determination of the correction factor for a cur-
rent input
Determination of the correction factor for, e.g. the measuring pressure input (menu
AB Absolute pressure
), which should be measured in a range of 20 to 70 bar.
1
st
step
Parameterize lower alarm limit parameter to 20 bar
(assigned to the technical zero point 0 or 4 mA).
2
nd
step
Parameterize upper alarm limit parameter to 70 bar
(assigned to the technical end value 20 mA).
3
rd
step
Parameterize the offset correction parameter to 0
4
th
step
Apply a pressure signal, and/or test current input with calibrated
measuring device and read the measured variable (display of meas-
ured pressure input in bar)
5
th
step
Determine the difference from:
actually supplied measuring signal and displayed measured variable
6
th
step
Enter this difference as an offset in the offset correction parameter
7
th
step
Check the pressure measurement display
The same procedure applies for all analog inputs.
Summary of Contents for ERZ2000-NG
Page 10: ......
Page 68: ...2 Introduction Manual ERZ 2000 NG EN07 2019 May 13th 58 Figure 44 Menu LA Overview...
Page 69: ...2 Introduction Manual ERZ 2000 NG EN07 2019 May 13th 59 Figure 45 Menu LB Totalizer BM1...
Page 77: ...2 Introduction Manual ERZ 2000 NG EN07 2019 May 13th 67 Figure 51 Overview System sub menu...
Page 241: ...6 Flow meters Manual ERZ 2000 NG EN07 2019 May 13th 231 Figure 167 Interface module C34...
Page 340: ...8 Overview Coordinates Manual ERZ 2000 NG EN07 2019 May 13th 330 8 5 6 TSV export...
Page 453: ...Certificates Manual ERZ 2000 NG EN07 2019 May 13th 443...