2 Introduction
Manual ERZ 2000-NG · EN07 · 2019, May 13th
75
Figure 62: Menu FF Revision
In order to receive logical values with appropriate resolution, an adequate test time
must be provided. A few minutes are sufficient for the volume detection via the HF
inputs, because synchronization of the test function with the recording of the volume
frequency takes place. For "slow" inputs, such as interfaces with ENCO or ultrasonic
gas meters, the test time must be sufficiently long to minimize resolution error (1000
seconds). This also applies for the "On-the-fly calibration" function.
The coordinates of the menu in detail:
FF01 Revision status
Shows the current status of the function
(at rest / running)
FF02 Time stamp 1
Parameter for starting the test process (run-up start)
FF03 Time stamp 2
Parameter for stopping the run-up and starting the test
FF04 Time stamp 3
Parameter for stopping the test and starting the after-run
FF05 Time stamp 4
Parameter for stopping the after-run and the function
FF06 Revision run
Parameter for a relative default of the test time, corre-
sponding to the time between Time stamp 3 and 4
Summary of Contents for ERZ2000-NG
Page 10: ......
Page 68: ...2 Introduction Manual ERZ 2000 NG EN07 2019 May 13th 58 Figure 44 Menu LA Overview...
Page 69: ...2 Introduction Manual ERZ 2000 NG EN07 2019 May 13th 59 Figure 45 Menu LB Totalizer BM1...
Page 77: ...2 Introduction Manual ERZ 2000 NG EN07 2019 May 13th 67 Figure 51 Overview System sub menu...
Page 241: ...6 Flow meters Manual ERZ 2000 NG EN07 2019 May 13th 231 Figure 167 Interface module C34...
Page 340: ...8 Overview Coordinates Manual ERZ 2000 NG EN07 2019 May 13th 330 8 5 6 TSV export...
Page 453: ...Certificates Manual ERZ 2000 NG EN07 2019 May 13th 443...