Appendix
Manual ERZ 2000-NG · EN07 · 2019, May 13th
355
A.4
Linking control totalizer with pulse output
Example: Control totalizer 1 (volume pulse) with pulse output 2
Note
There is no special operating mode for activation of extra totalizers. They
are supplied via the corresponding contact inputs and are active when a
control evaluation is parameterized unequal to zero and pulses arrive at
the input.
NL Frequency input 1
X8
–
7, 8
Control counter 1
NL03 Input pulses 1
x
NL10 Weighting
= NL08 Integer part
(e.g. 0.01)
MG Pulse output 2
MG19 Transit
= Control counter 1
MG10 Assignm.meas.val.
= Transit
NL08 Integer part
/
MG13 Pulse value
=
MG01 Pulse counter
(e.g. 2)
10 pulses
1000 pulses
5 pulses
1000 pulses