7 Parameter of the gas
Manual ERZ 2000-NG · EN07 · 2019, May 13th
291
7.4.9
DG Correction of velocity of sound
Figure 215: Menu DG Correction of velocity of sound
The ERZ2000-NG offers a correction option for direct density meters when the veloci-
ty of sound in the gas is known. For this purpose, the actual velocity of sound (also
abbreviated her as c
n
; c = velocity of sound under n = base conditions) must be en-
tered in coordinate
DG08 cn calib.gas
. It is compared with the correct velocity of
sound in
DG12 VOS for corr.
. With the formula for an ideal gas
ܿ ൌ ඨߢ
ߩ
Where
ܿ
-
Velocity of sound
ߢ
-
Adiabatic exponent
-
Presure
ߩ
-
Density
a correction value can be determined for the density (
DG02 Corr.factor Rho
). With
the path length along which the velocity of sound is determined, the change from
base to measurement conditions can be factored in.