3 Electrical connections
Manual ERZ 2000-NG · EN07 · 2019, May 13th
121
Resistance measurement
Type
PT 100 four-wire
Range
-20°C to +60°C
Resolution
0.01°C
Accuracy
0.05°C
Measuring time
50 ms
x
4 frequency inputs
Measuring range
0.00 Hz to 6.0 kHz
Accuracy
0.01 Hz
U hys
1.0 V
U trg
3.0 V
Overvoltage protection
6.8 V with external module
18.0 V with internal module (galvanically isolated)
x
8 signal inputs for H/L gas switching, travel direction reversal and external
freeze
Digital status inputs
All inputs are galvanically isolated from the computer, but not from each other. The
following can be used as signal transmitters: contact, open collector / drain, active
push / pull
-U max
5 V
-I max
13 mA
f max
10 Hz
Overvoltage protection
6.8 V
x
Reserve signal inputs
The ERZ2000-NG offers additional free inputs for which the same data as for the
"normal" signal inputs applies. These free inputs can be assigned with functions and
events, statuses, additional counter, etc. can be detected and stored in DSfG ar-
chives.
Appendix E) "Various circuit diagrams for inputs"
provides some input circuit
diagrams
Summary of Contents for ERZ2000-NG
Page 10: ......
Page 68: ...2 Introduction Manual ERZ 2000 NG EN07 2019 May 13th 58 Figure 44 Menu LA Overview...
Page 69: ...2 Introduction Manual ERZ 2000 NG EN07 2019 May 13th 59 Figure 45 Menu LB Totalizer BM1...
Page 77: ...2 Introduction Manual ERZ 2000 NG EN07 2019 May 13th 67 Figure 51 Overview System sub menu...
Page 241: ...6 Flow meters Manual ERZ 2000 NG EN07 2019 May 13th 231 Figure 167 Interface module C34...
Page 340: ...8 Overview Coordinates Manual ERZ 2000 NG EN07 2019 May 13th 330 8 5 6 TSV export...
Page 453: ...Certificates Manual ERZ 2000 NG EN07 2019 May 13th 443...