340
Testing and Calibrating the Digital Level
3
Place the level so that the display is facing you and lettering on the face of the level is right-side up.
Mark the position of the level so you can place the level back in the same spot when you move it during
this calibration process. Wait 10 seconds for the level to completely settle. Without moving the level,
gently press the
HOLD
button until
[ 1 ]
appears on the display.
4
Rotate the level so that it faces away from you and the screw on the back of the level is on the left side.
Ensure that the level is positioned in the same spot as
. Wait 10 seconds for the level to
completely settle. Without moving the level, gently press the
HOLD
button until
[ 2 ]
appears on the
display.
5
Roll the level towards you so the display is facing you but the lettering on the face of the level is upside
down. Ensure that the level is positioned in the same spot as
. Wait 10 seconds for the level to
completely settle. Without moving the level, gently press the
HOLD
button until an upside down
[ 3 ]
appears on the display.
Summary of Contents for Scalar i6000
Page 20: ...8 About This Guide and Your Product...
Page 38: ...26 Installing a Stand Alone Control Module...
Page 104: ...92 Installing a Multi Module Library...
Page 156: ...144 Installing Cartridges...
Page 164: ...152 Setting up Your Library for Access...
Page 242: ...230 Configuring the Library...
Page 304: ...292 Adding Optional Hardware eight 2 5 mm screws...
Page 318: ...306 Adding Optional Hardware...
Page 336: ...324 Installation Testing and Verification Figure 35 Example Test Log Output...
Page 356: ...344 Testing and Calibrating the Digital Level...
Page 362: ...350 LBX Board and Terminator...
Page 380: ...368 Glossary...
Page 384: ...372 Index...