AFX SERIES® OPERATION MANUAL
SECTION 8:
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 375 of 485
8.10.1.6
SYSTem:DIO:STROBE
Command Syntax
SYSTem:DIO:STROBE:OUTPustate <b>
Description
Sets the function strobe mode active when the output relay changes
state
Parameters
< 0 | OFF | 1 | ON >
Parameter Format
<b>
Example
SYST:DIO:STROBE:OUTP 1
Query Format
SYSTem:DIO:STROBE:OUTPustate?
Returned Data Format
<b>
Query Example
SYST:DIO:STROBE:OUTP?
1
Command Syntax
SYSTem:DIO:STROBE:SOURce <b>
Description
Sets the function strobe mode to program changes.
Parameters
< 0 | OFF | 1 | ON >
Parameter Format
<b>
Example
SYST:DIO:STROBE:SOUR 1
Query Format
SYSTem:DIO:STROBE:SOURce?
Returned Data Format
<b>
Query Example
SYST:DIO:STROBE:SOUR?
1
Command Syntax
SYSTem:DIO:STROBE:TRANsient<b>
Description
Sets the function strobe mode to generate an output at the start of a
transient execution.
Parameters
< 0 | OFF | 1 | ON >
Parameter Format
<b>
Example
SYST:DIO:STROBE:TRAN 1
Query Format
SYSTem:DIO:STROBE:TRANsient?
Returned Data Format
<b>
Query Example
SYST:DIO:STROBE:TRAN?
1
Summary of Contents for AFX Series
Page 484: ......