AFX SERIES® OPERATION MANUAL
SECTION 8:
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 302 of 485
The impact of inserting the initial value (ON) or not (OFF) is illustrated in the figure below.
•
The top sequence shows a STEP transient repeated 3 times with STEP:INIT set ot
OFF. The output will not return to the initial output setting between repetitions.
•
The bottom sequence is the same STEP transient with STEP:INIT set to ON. Each
repeat will start from the initial set value so each repititon produces the same
output levels.
Command Syntax
[SOURce:]STEP:DWELL <time>, <opt>
Description
This command sets the dwell time of the step in seconds
Minimum time set value allowed is 0.0002 sec. (0.2 msec)
Parameters
Time in seconds, Option: < NORmal | SATurate >
Default is NORmal if optional second parameter is omitted.
Parameter Format
<nr2>, <cr>
Example
STEP:MODE:DWELL 10, SAT
Query Format
[SOURce:]STEP:MODE?
Returned Data Format
<nr2>
Query Example
STEP:MODE:DWELL?
10.0000, SAT
Query Syntax
[SOURce:]STEP:DWELL:MINimum?
Description
This command returns lowest permissible set value for the STEP
dwell time setting.
Returned Data Format
<nr2>
Query Example
STEP:DWELL:MIN?
Query return values for MINimim, MAXimum and DEFault are dependent on
MODE and interdependent parameter value settings