AFX SERIES® OPERATION MANUAL
SECTION 13:
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 463 of 485
Error Source
Code
Category Error Message
Possible Cause
Resolution
with Output Control
Switch (OCS) option
which always enables
Remote Inhibit
Front panel
Error
126
Error
Line to line
measurements only
available with sinewaves.
Measurement not
available. Line to Line
voltage measurements
can be calculated from L-
N and Phase Angles only
when all phase
waveforms are Sine
Set all phases to use SINE
waveform #1.
Front panel
Error
127
Error
Digital I/O interface
conflict: Single-Phase
relay control is active.
Setting conflict
Front panel
Error
128
Error
Digital I/O interface
conflict: Single-Phase
relay cannot change with
output enable.
Setting conflict. Changing
Phase mode only allowed
when Output is OFF
Turn Output OFF
Front panel
Error
129
Error
Digital I/O interface
conflict: Single-Phase
relay cannot change due
relay closed.
Setting conflict
Front panel
Error
130
Error
Unable to disable output.
One or more inverter
stages failed to disable.
Output OFF failed as one
of more inverters could
not be disabled
Turn Unit off with Circuit
Breaker
Front panel
Error
144
Error
Internal battery needs to
be replaced.
Warning. Front panel
controller battery low.
Error could be result of
reverting to older
firmware version. If so,
use SANITIZE command
to clear this error.
Front panel
Error
148
Error
Serial number mismatch:
Front panel does not
match power stages.
Setting conflict. Front
panel does not match
power stages serial
number
Contact customer service
Front panel
Error
150
Error
Suffix not allowed in
single form.
Setting conflict.
Command suffix invalid
for Single Phase mode
Change phase mode
Front panel
Error
151
Error
Suffix not allowed in split
form.
Setting conflict.
Command suffix invalid
for Split Phase mode
Change phase mode
Front panel
Error
152
Error
Peak current limit setting
is too low for the desired
voltage and frequency.
Setting conflict. Peak
current limit setting too
low to support requested
voltage and frequency.
AFX has a minimum peak
limit because it affects
output inductor current
and at high frequencies, a
circulating current
proportional to the
voltage and frequency
occurs.
Increase the peak current
limit setting
Front panel
Error
153
Error
Program transient trigger
input is disabled.
Setting conflict. External
Trigger Input disabled
Enable External Trigger
Input
Front panel
Error
154
Error
Steady state not yet
stabilized.
??
??
Summary of Contents for AFX Series
Page 484: ......