AFX SERIES® OPERATION MANUAL
SECTION 8:
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 385 of 485
8.11
IEEE488.2 Common Commands
The following IEEE488.2 common commands (a.k.a. star commands) are supported by the
AC power source. These commands are provided for compatibility with the IEEE488.2
standard. They are aliases to the relevant proprietary command and can be used
interchangeably. Commands are shown in alphabetical order.
IEEE488.2 Command Description
Group
Mandatory
*CLS
Clear Status
Status and Event
Yes
*DCL
Device Clear
Internal Operations
*ESE <n>
Event Status Register Enable
Status and Event
Yes
*ESE?
ESE Query
Status and Event
Yes
*ESR?
Event Status Register Query
Status and Event
Yes
*GTL
Goto Local
Control
*IDN?
Identify
System Data
Yes
*LLO
Local Lock Out
Control
Yes
*OPC
Operation Complete
Synchronization
Yes
*OPC?
OPC Status Query
Synchronization
Yes
*RST
Reset
Internal Operations
Yes
*SRE
Service Request Enable
Status and Event
Yes
*SRE?
SRE Query
Status and Event
Yes
*STB?
Status Byte Query
Status and Event
Yes
*TRG
Trigger
Synchronization
*WAI
Wait
Synchronization
Yes
Table 8-3: Mandatory IEEE488.2 Common Commands
Command Syntax
*CLS
Description
Clear Status. The Clear Status (CLS) command clears the status byte
by emptying the error queue and clearing all the event registers
including the Data Questionable Event Register, the Standard Event
Status Register, the Standard Operation Status Register and any
other registers that are summarized in the status byte.
Command Syntax
*DCL
Description
Device Clear. Resets the instrument to a default state.
Parameters
None
Parameter Format
n/a