Command Syntax
[SOURce:]COUPLing <cr>
Description
Selects the output coupling mode. On AFX Series®, this command
only accepts DIRECT as a setting unless the optional output
transformer is connected and configured. The output voltage ratio of
the transformer can be queried using the “SYSTem:XFMRRATIO?”
command. The output voltage range is 300Vac x Ratio full scale Line
to Neutral.
If no transformer option is installed and configured, selecting XMFR
coupling will generate a 2019 Error code. Do not select XMFR
coupling if the SYSTem:XFMRRATIO? query returns a 0.0000 value.
Parameters
< DIRECT | 0 | XMFR | 1 >
0 = DIRECT
1 = XMFR
Parameter Format
<cr>
Example
COUPL DIRECT
Query Format
[SOURce:]COUPLing?
Returned Data Format
<cr>
Query Example
COUPL?
0