AFX SERIES® OPERATION MANUAL
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 146 of 485
6.6.11
PULSE Parameters
Pulse transients can be used to create repetitive events for endurance testing of AC and DC
powered products.
The following parameters are available in a PULSE VALUES screen.
PARAMETER
Range
Unit
Description
Frequency
15 ~ 1200
Hz
Frequency
Voltage AC
0 ~ 300
Vrms
AC Voltage
Voltage DC
-425 ~ + 425
Vdc
DC Voltage
Waveform
1 ~ 200
Waveform number
Phase
0.0 ~ 359.9
Deg
Phase angle for phases B & C
Table 6-17: Available STEP Transient Parameters
The following soft keys are available on the PULSE VALUES screen.
SOFT KEY
Description
Use Setpoint (SK1)
Starts the transient from the “Run from step#” row number
Run Screen(SK2)
Displays the PULSE EXECUTION screen.
Link Phases (SK3)
Toggles between phase A, B, C or Coupled ABC
Example (SK4)
Displays a graphical representation of the STEP parameters
Back (SK5)
Returns to previous screen
Table 6-18: Available STEP PROGRAM screen soft keys
Summary of Contents for AFX Series
Page 484: ......