Command Syntax
*OPC
Description
The Operation Complete (OPC) command sets bit 0 in the Standard
Event Status Register when all pending operations have finished.
Parameters
Optional: < BLOCK | 0 | NOBLOCK | 1>
Example
*OPC
Query Format
*OPC? < BLOCK | 0 | NOBLOCK | 1>
Description
IEEE488.2 standard command. The parameter is optional.
The argument is optional, if it is not sent:
•
In UPC compatible mode default argument will be NONBLOCK or
1.
•
In normal mode default argument will be BLOCK or 0.
*OPC? BLOCK | 0
Returns 1 when all pending overlapped operations have been
completed. It can be used to cause the controller to wait for
commands to complete.
*OPC? NOBLOCK | 1
Returns 1 if all pending overlapped operations have been completed
or 0 if there are pending overlapped operations. It will
not
cause the
controller to wait for commands to complete. Pending overlapped
operations can be a transient or a soft start using ramp time/slew
rates.
Returned Data Format
<b>
Query Example
*OPC?
1
Command Syntax
*RST
Description
RESET. The *RST command (reset) has the same effect as an IEEE-488
Device Clear bus command but can be used over the RS232C, USB or
LAN interface as well. This command resets the unit to its power on
default state.
Parameters
None
Parameter Format
n/a
RESET STATE
FORM
3
VOLT:MODE
AC
VOLT:AC
0.0000
COUPLING
DC
VOLT:DC
0.0000
RANGE
AC
CURR:AC
41.6667
CURR:DC
20.8333