8. Troubleshooting
PROGRAMMABLE AC/DC POWER SOURCE
218
Table 8-3 Actions When a Failure is Suspected (Continued)
Category Symptom
Possible
Cause Necessary
Actions
Sequence/
Simulation-
related
problem
The Sequence
cannot be
started.
The output is off.
Turn on the output, then start the
sequence.
The specified
Step Time is
exceeded.
The Stop Phase is
enabled.
When the stop phase setting is
active, the execution waits for the
Step Time that was set pass, and
keeps the output setting until it
reaches the Stop Phase that was
set, and then transitions to the
next step. As the result, the actual
step time is longer than the set
Step Time. To move to the next
step immediately after the Step
Time, disable the Stop Phase.