8.1 Error Messages and Error Handling
KP3000GS
211
Table 8-2 Error Message List (Continued)
ID Message Protection
type
Cause or Necessary action
50
Output Overcurrent [RMS] Output Off
The output current RMS value is excessive
on a module in the power unit. This may
occur when the output terminal is
short-circuited.
51
Output Overcurrent [Peak] Output Off
The output current peak value is excessive.
52
Power Unit DCPS Error
Output Off
The DC power supply in the power unit is
abnormal, or see the notes in "
8.1.3
,
Operation procedure."
53
Power Unit DCPS
Overvoltage
Output Off
The voltage of the DC power supply in the
power unit is excessive. This may occur
when the power flows back from the load.
54
Power Unit DCPS
Undervoltage
Output Off
The voltage of the DC power supply in the
power unit is insufficient, or see the notes in
"
8.1.3
, Operation procedure." This may
occur when the overload output activated
the protection function on the DC power
supply.
55
Overheat
Output Off
The temperature in the power unit is
abnormally high. Check if the ambient
temperature is appropriate. Check the filters
for clogging. For filter cleaning, see
9.2
.
56
Sensing Voltage Error
Output Off
The sensing voltage is abnormal. Make sure
the cable is properly connected to the
sensing input terminal.
57
Sync Frequency Error
Output Off
The frequency went out of the
synchronizable range, and thus the output
turned off.
58
Current Limiter [RMS]
Output Off
The RMS current limiter was activated, and
thus the output was turned off according to
the setting.
59
Current Limiter [Peak]
Output Off
The peak current limiter was activated, and
thus the output was turned off according to
the setting.
60
Power Unit Internal Error
Output Off
This is a power unit internal error. If it
occurs repeatedly, contact the seller (NF
Corporation or our agent) from whom you
purchased the product.
61
Limiting
action
The active wattage limiter is activated.
62
Limiting
action
The RMS current limiter is activated.
63
Limiting
action
The peak current limiter is activated.
The internal operation differs from that for
ID 64.
64
Limiting
action
The peak current limiter is activated.
The internal operation differs from that for
ID 63.
68 PU-ON
Setting
Updated Message
display only
The power unit energization setting was
updated.
Summary of Contents for KP3000GS
Page 1: ...PROGRAMMABLE AC DC POWER SOURCE KP3000GS INSTRUCTION MANUAL...
Page 2: ......
Page 3: ...DA00033845 005 PROGRAMMABLE AC DC POWER SOURCE KP3000GS INSTRUCTION MANUAL...
Page 4: ......
Page 5: ......
Page 19: ......
Page 20: ...KP3000GS 1 1 Outline 1 1 Overview 2 1 2 Features 2...
Page 23: ......
Page 197: ......
Page 207: ......
Page 219: ......
Page 243: ......
Page 275: ......
Page 277: ......
Page 278: ......