Table 8-2 Error Message List (Continued)
ID Message Protection
type
Cause or Necessary action
32
Calibration Data Error ME Message
display only
The adjustment data is abnormal. Cycle the
power. If it still occurs, contact the seller
(NF Corporation or our agent) from whom
you purchased the product.
33 Line
Overvoltage
System
Lock
The power input voltage is excessive. Check
if the power input is appropriate.
34 Line
Undervoltage
System
Lock
The power input voltage is insufficient.
Check if the power input is appropriate.
36 Communication
Failure
PU
System
Lock
There is a communication error between the
firmware and the power unit. Cycle the
power. If it still occurs, contact the seller
(NF Corporation or our agent) from whom
you purchased the product.
37 Communication
Failure
ME
System
Lock
There is a communication error between the
firmware and the output measurement
section. Cycle the power. If it still occurs,
contact the seller (NF Corporation or our
agent) from whom you purchased the
product.
43
Mismatched Pw Unit
Versions
System
Lock
All power unit versions do not match.
44
Unsupported Pw Unit
Version
System
Lock
A power unit of unsupported version is
connected.
47
No Available Power Unit
System
Lock
There is no power unit that can be started.
48
Output Overvoltage
Output Off
The output voltage is excessive. This may
occur at an abrupt change in the output
current due to the inductive load.
Summary of Contents for KP3000GS
Page 1: ...PROGRAMMABLE AC DC POWER SOURCE KP3000GS INSTRUCTION MANUAL...
Page 2: ......
Page 3: ...DA00033845 005 PROGRAMMABLE AC DC POWER SOURCE KP3000GS INSTRUCTION MANUAL...
Page 4: ......
Page 5: ......
Page 19: ......
Page 20: ...KP3000GS 1 1 Outline 1 1 Overview 2 1 2 Features 2...
Page 23: ......
Page 197: ......
Page 207: ......
Page 219: ......
Page 243: ......
Page 275: ......
Page 277: ......
Page 278: ......