3.4 Using the Continuous Function
KP3000GS
41
--------
Notes
------------------------------------------------------------------------------------------------------------
Set the RMS values for the voltages of the sine wave (SIN) and the clipped sine wave (CLP)
and the Peak-to-Peak value for the voltage of the arbitrary waveform (ARB).
-----------------------------------------------------------------------------------------------------------------------------
Operation procedures for setting the DC voltage
1.
Select the [DCV] item and then enter a numeric value. For how to enter values, see
3.3.5
.
3.4.5 Setting the Output Frequency
Set the AC component frequency of the internal signal source.
Operation procedure
1.
Select the [Freq] item and follow the instructions in
3.3.5
to enter a value.
3.4.6 Setting the Output On/Off Phase
Set the phases when the output turns on/off. After the output-on operation, the voltage output is
started from the output on phase. After the output off operation, the voltage output is completed
when it reaches the output off phase. The output off phase can be disabled. If disabled, the output is
turned off immediately after the operation of output off.