106
OPERATOR'S MANUAL
EN
15.2. TECHNICAL MAINTENANCE
These instructions describe the maintenance procedures for the RX DC X-ray unit. These instructions apply to all the revisions of said equipment.
In order to ensure the operational safety and functional reliability of the equipment installed, at least once a year an authorized service technician must perform a full inspection of the device.
When taking measurements that require a multimeter, always use a calibrated digital multimeter.
All the following tests will be carried out. Customer should be notified prior to replacing parts.
Answer questions with yes (√) or no (−)
Inspection DATE
Step
Description
Reference in User Manual
__/__/20__
__/__/20__
__/__/20__
__/__/20__
__/__/20__
1
Check that all labels located
- on the wall-mounted cover,
- on the X-ray tube
- inside the collimator/collimators,
are intact, well attached and legible.
User Manual, section Identification nameplates
2
Check there are no external damages to the equipment which could that may
reduce protection against radiation.
User Manual, section Description of the RX DC x-ray unit
3
Pull out the collimator and panel stop ring, take off the screw cover caps and
loosen the screws that secure the lower cover. Check there is no oil leakage on
the tube-head.
Technical Manual, section X-ray Head
4
Check the electromechanical brake that locks/unlocks the movement of the
generator and adjust it if necessary – Hypersphere model units only –
Technical Manual, section Actuator unit
5
Power off the unit and remove the wall mounting cover. Disconnect the unit
from the main power supply and check the condition of the main power supply
cable. Replace it in case of damage. Connect it back making sure the safety
ground is securely connected. Install the wall mounting cover back again.
Technical Manual, section Wall mounted plate wiring connections
6
Check the battery level of the wireless remote control
User Manual, section Batteries and charge level indication
7
Check the remote control functionality: buttons must respond to interaction
User Manual, section Control Panel
8
Check the power switch verifying that the switch is working properly and the
main switch green light illuminates when the switch is in the on position.
Section Turning the x-ray unit on and off
9
Check proper functioning of X-ray generator indicator light - Hypersphere model
units only
User Manual, section X-ray generator indicator light
10
Check the exposure buzzer during a trial x-rays emission
User Manual, section Control Panel
11
Verify that exposure is immediately interrupted when X-rays button is released
Section Control Panel
12
Check the scissors arm balance and adjust it if necessary
Technical Manual, section Balancing the double pantograph arm
13
Perform a calibration of the X-ray tube housing
Technical Manual, section Calibrating the X-ray Head
Operator Name
Signature
The undersigned confirms that he/she has checked the unit for the above criteria and that he/she has handed over the unit in fully functional condition.
All inspection and maintenance work performed by the system owner and/or service engineer must be recorded in this document and kept near the unit!.
Summary of Contents for RX DC
Page 1: ...97050850 Rev 12 15 10 myray RX DC RX DC I IT EN FR ES PL RU ...
Page 2: ...2 ISTRUZIONI PER L USO IT ITALIANO ...
Page 4: ...4 ISTRUZIONI PER L USO IT ...
Page 29: ...IT ISTRUZIONI PER L USO 29 ...
Page 31: ...IT ISTRUZIONI PER L USO 31 ...
Page 34: ...34 ISTRUZIONI PER L USO IT ...
Page 52: ...52 ISTRUZIONI PER L USO IT ...
Page 56: ...56 OPERATOR S MANUAL EN ...
Page 81: ...EN OPERATOR S MANUAL 81 ...
Page 83: ...EN OPERATOR S MANUAL 83 ...
Page 85: ...EN OPERATOR S MANUAL 85 11 3 XT VERSION All dimensions are expressed in millimeters inches ...
Page 86: ...86 OPERATOR S MANUAL EN ...
Page 104: ...104 OPERATOR S MANUAL EN ...
Page 108: ...108 INSTRUCTIONS D UTILISATION FR ...
Page 133: ...FR INSTRUCTIONS D UTILISATION 133 ...
Page 137: ...FR INSTRUCTIONS D UTILISATION 137 ...
Page 156: ...156 INSTRUCTIONS D UTILISATION FR ...
Page 160: ...160 INSTRUCCIONES DE USO ES ...
Page 187: ...ES INSTRUCCIONES DE USO 187 ...
Page 206: ...206 INSTRUCCIONES DE USO ES ...
Page 210: ...210 INSTRUKCJA OBSŁUGI PL ...
Page 236: ...236 INSTRUKCJA OBSŁUGI PL 11 3 WERSJA XT Wszystkie wymiary wyrażone są w milimetrach calach ...
Page 237: ...PL INSTRUKCJA OBSŁUGI 237 ...
Page 256: ...256 INSTRUKCJA OBSŁUGI PL ...
Page 260: ...260 ИНСТРУКЦИИ ПО ПРИМЕНЕНИЮ RU ...
Page 286: ...286 ИНСТРУКЦИИ ПО ПРИМЕНЕНИЮ RU 11 3 ВАРИАНТ XT Все размеры выражены в миллиметрах дюймах ...
Page 287: ...RU ИНСТРУКЦИИ ПО ПРИМЕНЕНИЮ 287 ...
Page 306: ...306 ИНСТРУКЦИИ ПО ПРИМЕНЕНИЮ RU ...
Page 309: ......