64
OPERATOR'S MANUAL
EN
3.3. CONTROL PAD
As illustrated in the figure below, the handheld has four function keys and a single x-ray emission key.
1
“Dentition area selection” key
2
“Body build selection” key
3
X-ray emission light
4
"Increase" key
5
"Decrease" key
6
"X-ray emission light" key
The main functions of the keys on the handheld vary according to how they are pressed:
KEY
BRIEFLY PRESSED (less than 3 sec.).
PRESSED LONGER (more than 3 sec.).
Changes over from ADULT to CHILD and
vice versa (takes place when key is
released).
Saves the selected setting (exposure time,
sensitivity, etc…). The memo icon
(
) lights up
when the data item can be saved.
It will change over the various types of teeth
to select the area to be examined.
It displays the values corresponding to the tooth
exposure times in mGy and in mGy*cm
2
if
pressed again.
Increases the exposure times in steps,
according to the set scale.
Increases the scroll speed of the values in
increasing order.
Decreases the exposure times in steps,
according to the set scale.
Increases the scroll speed of the values in
decreasing order.
NO EFFECTS ARE OBTAINED IF THE
KEY IS PRESSED LESS THAN A
SECOND.
Starts x-ray exposure (the button has to be held
down while the x-rays are being emitted, “dead
man” function).
NOTE: “Dead man” function: the system that starts x-ray exposure with the dedicated key on the wireless
handheld allows x-rays to be emitted only when the user presses and holds down the exposure key. X-ray
emission will stop if the key is released ahead of time.
Summary of Contents for RX DC
Page 1: ...97050850 Rev 12 15 10 myray RX DC RX DC I IT EN FR ES PL RU ...
Page 2: ...2 ISTRUZIONI PER L USO IT ITALIANO ...
Page 4: ...4 ISTRUZIONI PER L USO IT ...
Page 29: ...IT ISTRUZIONI PER L USO 29 ...
Page 31: ...IT ISTRUZIONI PER L USO 31 ...
Page 34: ...34 ISTRUZIONI PER L USO IT ...
Page 52: ...52 ISTRUZIONI PER L USO IT ...
Page 56: ...56 OPERATOR S MANUAL EN ...
Page 81: ...EN OPERATOR S MANUAL 81 ...
Page 83: ...EN OPERATOR S MANUAL 83 ...
Page 85: ...EN OPERATOR S MANUAL 85 11 3 XT VERSION All dimensions are expressed in millimeters inches ...
Page 86: ...86 OPERATOR S MANUAL EN ...
Page 104: ...104 OPERATOR S MANUAL EN ...
Page 108: ...108 INSTRUCTIONS D UTILISATION FR ...
Page 133: ...FR INSTRUCTIONS D UTILISATION 133 ...
Page 137: ...FR INSTRUCTIONS D UTILISATION 137 ...
Page 156: ...156 INSTRUCTIONS D UTILISATION FR ...
Page 160: ...160 INSTRUCCIONES DE USO ES ...
Page 187: ...ES INSTRUCCIONES DE USO 187 ...
Page 206: ...206 INSTRUCCIONES DE USO ES ...
Page 210: ...210 INSTRUKCJA OBSŁUGI PL ...
Page 236: ...236 INSTRUKCJA OBSŁUGI PL 11 3 WERSJA XT Wszystkie wymiary wyrażone są w milimetrach calach ...
Page 237: ...PL INSTRUKCJA OBSŁUGI 237 ...
Page 256: ...256 INSTRUKCJA OBSŁUGI PL ...
Page 260: ...260 ИНСТРУКЦИИ ПО ПРИМЕНЕНИЮ RU ...
Page 286: ...286 ИНСТРУКЦИИ ПО ПРИМЕНЕНИЮ RU 11 3 ВАРИАНТ XT Все размеры выражены в миллиметрах дюймах ...
Page 287: ...RU ИНСТРУКЦИИ ПО ПРИМЕНЕНИЮ 287 ...
Page 306: ...306 ИНСТРУКЦИИ ПО ПРИМЕНЕНИЮ RU ...
Page 309: ......