70
OPERATOR'S MANUAL
EN
The exposure time can also be modified by changing the sensitivity factor. Press keys
and
at the same
time , the actual sensitivity factor will be displayed.
Use keys
and
to change the value from 3 to 25. If the displayed value differs from the default setting, icon
comes on. To quit this mode, press key
or
. The change made to the sensitivity factor, is applied to all
the teeth and both patient sizes.
The selected operating mode is always used for each tooth and patient size combination in modes En60, En63 and
En65.
In AUTO mode, each tooth and patient size combination is associated to the best mode from amongst the ones
available. In this mode it is not possible to assign a mode other than the default one to each combination.
To set the mode, refer to paragraph 4.5 “Setting the mode and exposure time in USER mode”.
To change the mode amongst En60, En63, En65 and AUTO refer to paragraph 5.2 “Setting the operating mode”.
4.5. SETTING THE MODE AND EXPOSURE TIME IN USER MODE
In USER mode, it is possible to assign an exposure time and a mode from amongst En60, En63 and En65 to each
tooth-patient size combination.
The default setting corresponds to the AUTO mode settings with sensitivity factor F=19.
To activate USER mode regardless of the mode currently being used, press keys
and
at the same time.
Icon
will come on to signal USER mode is active.
To deactivate USER mode press keys
and
again (icon
goes off).
The exposure parameters are set by following the directions given below::
1) Select the tooth to be examined
2) Select the patient size
The exposure time is automatically shown on the handheld screen.
NOTE: It is not possible to access the sensitivity factor menu in USER mode. In addition, keys
and
are inoperative in User mode.
The exposure times and mode assigned to the tooth – patient size combinations are custom set by following the
directions given below:
1) Press and hold down key
about two seconds. Customized settings can be entered and icon
comes on.
2) Select the desired tooth-patient size combination.
3) The exposure time can be changed with keys
and
.
NOTE: Exposure times ranging from 0.01s and 1.00s belonging to the R’20 scale can be set.
Summary of Contents for RX DC
Page 1: ...97050850 Rev 12 15 10 myray RX DC RX DC I IT EN FR ES PL RU ...
Page 2: ...2 ISTRUZIONI PER L USO IT ITALIANO ...
Page 4: ...4 ISTRUZIONI PER L USO IT ...
Page 29: ...IT ISTRUZIONI PER L USO 29 ...
Page 31: ...IT ISTRUZIONI PER L USO 31 ...
Page 34: ...34 ISTRUZIONI PER L USO IT ...
Page 52: ...52 ISTRUZIONI PER L USO IT ...
Page 56: ...56 OPERATOR S MANUAL EN ...
Page 81: ...EN OPERATOR S MANUAL 81 ...
Page 83: ...EN OPERATOR S MANUAL 83 ...
Page 85: ...EN OPERATOR S MANUAL 85 11 3 XT VERSION All dimensions are expressed in millimeters inches ...
Page 86: ...86 OPERATOR S MANUAL EN ...
Page 104: ...104 OPERATOR S MANUAL EN ...
Page 108: ...108 INSTRUCTIONS D UTILISATION FR ...
Page 133: ...FR INSTRUCTIONS D UTILISATION 133 ...
Page 137: ...FR INSTRUCTIONS D UTILISATION 137 ...
Page 156: ...156 INSTRUCTIONS D UTILISATION FR ...
Page 160: ...160 INSTRUCCIONES DE USO ES ...
Page 187: ...ES INSTRUCCIONES DE USO 187 ...
Page 206: ...206 INSTRUCCIONES DE USO ES ...
Page 210: ...210 INSTRUKCJA OBSŁUGI PL ...
Page 236: ...236 INSTRUKCJA OBSŁUGI PL 11 3 WERSJA XT Wszystkie wymiary wyrażone są w milimetrach calach ...
Page 237: ...PL INSTRUKCJA OBSŁUGI 237 ...
Page 256: ...256 INSTRUKCJA OBSŁUGI PL ...
Page 260: ...260 ИНСТРУКЦИИ ПО ПРИМЕНЕНИЮ RU ...
Page 286: ...286 ИНСТРУКЦИИ ПО ПРИМЕНЕНИЮ RU 11 3 ВАРИАНТ XT Все размеры выражены в миллиметрах дюймах ...
Page 287: ...RU ИНСТРУКЦИИ ПО ПРИМЕНЕНИЮ 287 ...
Page 306: ...306 ИНСТРУКЦИИ ПО ПРИМЕНЕНИЮ RU ...
Page 309: ......