MARKer Subsystem
11
Keysight 8990B Programming Guide
327
Example
Remarks
– If a disabled channel is specified as the marker source, error –221,“Settings
conflict; Requires channel # to be enabled” will occur.
– This command is only applicable for the marker spacing measurement. If this
command is sent for a marker measurement mode other than spacing, error
–221,“Settings conflict; Requires to switch to pulse spacing marker
measurement” will occur.
– The marker spacing measurement only works when the marker source is set to
the same channel source. Use this command to set the automatic spacing
marker instead of the
MARKer:ASOUrce
and
MARKer:BSOUrce
command.
Reset condition
On reset, the marker measurement mode is set to
OFF
.
Query
MARKer:SPACing:SOURce?
This query returns the current setting of the marker source for spacing
measurement.
Query example
MARK:SPAC:SOUR CHAN1
This command sets the marker source for spacing
measurement to channel 1.
MARK:SPAC:SOUR?
Queries the marker source for spacing measurement.