9
MEASure Subsystem
260
Keysight 8990B Programming Guide
Example
Remarks
– Set the acquire mode to normal, zoom, or multipulse mode before using this
command. Refer to
“ACQuire:MODE <character_data>”
– Normal mode measurement queries are allowed in zoom mode.
– Zoom mode measurement queries are allowed only in zoom mode. If this query
is made when in other modes, error -221,“Settings conflict;Require zoom
mode to be enabled” will occur.
– Multipulse mode measurement queries are allowed only in multipulse mode. If
this query is made when in other modes, error -221,“Settings conflict;Require
multipulse mode to be enabled” will occur.
– The multipulse query is only applicable when the multipulse acquisition is
completed in the multipulse mode. If this query is sent when the multipulse
acquistion is running, error –221,“Stop multipulse frame acquisition to change
the setting” will occur.
MEAS:PWID? CHAN1
Queries the pulse width measurement result
for channel 1 in normal mode.
MEAS:PWID? CHAN1, ZOOM
Queries the pulse width measurement result
for channel 1 in zoom mode.
MEAS:PWID? CHAN1, MULT
Queries the pulse width measurement result
for channel 1 in multipulse mode.