19
MEMory Subsystem
490
Keysight 8990B Programming Guide
Remarks
– If no FDO table has been selected when sending this command, error
–221,“Settings conflict;No FDO table selected” will occur.
– If a frequency value is specified which is a duplicate of an existing frequency
value, error –221,“Settings conflict;Duplicated frequency input” will occur.
– If you set an offset value which exceeds its minimum or maximum limit, error
–222,“Data out of range” will occur.
– If the selected FDO table has more than 80 frequency points, error
946,”Selected FDO table has reached the limit of 80 frequency points” will
occur.