Example 2: Making Pulse Characterization Measurement
The following command sequences provide the example to perform the pulse
characterization measurement.
This measurement example requires the following test setup:
Figure 23-1
Test setup for the pulse characterization measurement
E4438C signal generator configuration
NOTE
“–>” indicates the commands that you send to the 8990B.
E4438C signal generator
8990B peak power analyzer
N1923/24A sensor
-> SYST:PRES
//Presets the E4438C to its default settings.
Wait 1 sec for operation to complete.
-> *CLS
//Clears the error queue.
-> FREQ 1GHZ
//Sets the frequency to 1 GHz.
-> POW 0DBM
//Sets the output power level to 0 dBm.
-> PULM:INT:PER 1000uS
//Sets the pulse period to 1000
μ
s.
-> PULM:INT:PWID 200uS
//Sets the pulse width to 200
μ
s.
-> PULM:STAT ON
//Turns on the pulse signal.