9
MEASure Subsystem
266
Keysight 8990B Programming Guide
Parameter
Example
Remark
If you set a reference level value which exceeds its minimum or maximum limit, the
value will be clipped to its minimum or maximum value respectively. Error
–222,“Data out of range;Value clipped to minimum (#)” or –222,“Data out of
range;Value clipped to maximum (#)” will occur.
Reset condition
On reset, the trace reference level is set to 10% for reference level 1, and 90% for
reference level 2.
Item
Description/Default
Range of values
numeric_value
A numeric value in % for the trace reference level:
– Default value:
– 10% (for reference level 1)
– 90% (for reference level 2)
– Minimum value: 0%
– Maximum value: 100%
– Reference level 1
– Maximum value: < reference level 2
– Reference level 1 - Maximum value < Pulse Duration
– Reference level 2
– Minimum value: > reference level 1
– Reference level 2 - Minimum value > Pulse Duration
0 to 100%
MEAS:THR:REF1 CHAN1,20
This command sets the trace reference level
1 to 20% for channel 1.