13
TRACe Subsystem
382
Keysight 8990B Programming Guide
Parameter
Example
Remarks
– For the channel source, this query is only applicable when the channel is
enabled. If the channel is disabled, error –221,“Settings conflict;Requires
channel # to be enabled” will occur.
– For the CCDF source, this query is only applicable when the CCDF acquisition
mode is enabled. If the CCDF acquisition mode is not enabled, error
–221,“Settings conflict;Requires CCDF mode to be enabled” will occur. Refer
“ACQuire:MODE <character_data>”
on page 132 to set the CCDF acquisition
mode.
– For the CCDF Gaussian source, this query is only applicable when the CCDF
Gaussian trace is enabled. If the Gaussian trace is not enabled, error
–221,“Settings conflict;Requires CCDF Gaussian trace to be enabled” will
occur. Refer to
“ACQuire:CCDF:GAUSsian[:STATe] <boolean>”
set the CCDF Gaussian trace.
Item
Description/Default
Range of values
character_data
Sets the source as:
–
CHAN1
: Sets the channel 1 trace.
–
CHAN2
: Sets the channel 2 trace.
–
CHAN3
: Sets the channel 3 trace.
–
CHAN4
: Sets the channel 4 trace.
–
CCDF1
: Sets the channel 1 CCDF trace.
–
CCDF4
: Sets the channel 4 CCDF trace.
–
CREFerence
: Sets the CCDF reference trace.
–
CGAUssian
: Sets the CCDF Gaussian trace.
CHAN1
CHAN2
CHAN3
CHAN4
CCDF1
CCDF4
CREFerence
CGAUssian
TRAC:DATA? CHAN1
Queries the current trace data of channel 1.
Summary of Contents for 8990B
Page 1: ...Keysight 8990B Peak Power Analyzer Programming Guide ...
Page 16: ...16 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...20 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...