ACQuire:CCDF:TRACe{1|4}[:STATe] <boolean>
This command sets the state of the CCDF trace from channel 1 or 4.
Syntax
Example
Remarks
– This command is only applicable for channels 1 and 4. If you send this
command for a channel other than 1 or 4, error 700,“Applicable to channel 1
and 4 only” will occur.
– This command is only applicable in the CCDF acquisition mode. If the
acquisition mode is not set to CCDF, error –221,“Settings conflict;Requires
CCDF mode to be enabled” will occur. Refer to
on page 132 to set the acquisition mode.
Reset condition
On reset, the CCDF trace from both channels 1 and 4 is turned on.
ACQ:CCDF:TRAC1 ON
This command turns on the CCDF trace from
channel 1.
ACQ
:STAT
:CCDF
0|OFF
1|ON