TRIGger Subsystem
8
Keysight 8990B Programming Guide
219
TRIGger:PWIDth <mode>,<time1>,<time2>
This command sets the mode and time for the pulse width trigger.
Syntax
Parameter
Example
Reset condition
On reset, the pulse width trigger mode is set to
MORE
and the time is set to 1
μ
s.
Item
Description/Default
Range of values
mode
Specifies the pulse width trigger mode as:
–
LESS
: Sets the pulse width mode to less. Only time1 is required.
–
MORE
: Sets the pulse width mode to more. This is the default
setting. Only time1 is required.
–
WITHIN
: Sets the pulse width mode to within. Time 1 and time 2
are required.
–
OUTOF
: Sets the pulse width mode to out. Time 1 and time2 are
required.
LESS
MORE
WITHIN
OUTOF
time1
time2
A numeric value for the pulse width.
Default:
1
μ
s
60 ns to 500 ms
TRIG
Space
?
:PWID
mode
time1
,
time2
,
TRIG:PWID LESS,0.1
This command sets the pulse width trigger mode
to less at 0.1 s.