ACQuire Subsystem
4
Keysight 8990B Programming Guide
91
Remarks
– This command is only applicable for channels 1 and 4. If you send this
command for a channel other than 1 or 4, error 700,“Applicable to channel 1
and 4 only” will occur.
– This command is only applicable in the CCDF acquisition mode. If the
acquisition mode is not set to CCDF, error –221,“Settings conflict;Requires
CCDF mode to be enabled” will occur. Refer to
on page 132 to set the acquisition mode.
– If you set a trigger duration value which exceeds its minimum or maximum
limit, the value will be clipped to its minimum or maximum value respectively.
Error –222,“Data out of range;Value clipped to minimum (#)” or –222,“Data out
of range;Value clipped to maximum (#)” will occur.
– This command is only applicable when the CCDF sweep is set to the triggered
mode. If the CCDF sweep mode is not set to triggered, error –221,“Settings
conflict; Requires CCDF sweep to be in triggered mode” will occur.
– CCDF duration is set based on the trace length of the normal mode when
switching from normal to zoom mode. This command is applicable within the
range of duration, If the duration is out of range,“Settings conflict;CCDF
trigger duration is out of range. It sets to default” will occur.
Reset condition
On reset, the CCDF trigger duration is set to 10 ms.
Query
ACQuire:CCDF:DURation?
This query returns the current setting of the CCDF trigger duration. The response
format is
<NRf>
.
Query example
ACQ:CCDF:DUR?
Queries the CCDF trigger duration setting.
Summary of Contents for 8990B
Page 1: ...Keysight 8990B Peak Power Analyzer Programming Guide ...
Page 16: ...16 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...20 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...