SERVice Subsystem
21
Keysight 8990B Programming Guide
515
SERVice:CSOUrce:INTerval <numeric_value>
This command sets the time interval for the reference check source.
Syntax
Parameter
Example
Remark
If you set a check source interval value which exceeds its minimum or maximum
limit, the value will be clipped to its minimum or maximum value respectively.
Error –222,“Data out of range;Value clipped to minimum (#)” or –222,“Data out of
range;Value clipped to maximum (#)” will occur.
Reset condition
On reset, the reference check source interval is set to 10 min.
SERV
:CSOU
Space
?
:INT
numeric_value
Item
Description/Default
Range of values
numeric_value
A numeric value in minute for the check source interval:
– Default value: 10 min
– Minimum value: 1 min
– Maximum value: 60 min
1 min to 60 min
SERV:CSOU:INT 20
This command sets the reference check source
interval to 20 min.
Summary of Contents for 8990B
Page 1: ...Keysight 8990B Peak Power Analyzer Programming Guide ...
Page 16: ...16 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...20 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...