Remark
If this query is sent when marker A and marker B have different measurement
levels, error –400,“Query error; Unable to query marker level. Marker A and
Marker B have different level. Refer to Programming Guide for further
information.” will occur. You may use
or
to query the measurement level.
Reset condition
On reset, the marker measurement level is set to 50%.
Query
MARKer:LEVel?
This query returns the current setting of the marker measurement level. The
response format is
<NRf>
.
Query example
MARK:LEV 55
This command sets the marker measurement level to 55%.
MARK:LEV?
Queries the marker measurement level setting.