CALibration:OUTPut <character_data>
This command sets the AC coupling frequency or DC level of the calibrator
waveform output through the rear panel Auxiliary Trigger Out connector.
Syntax
Parameter
Example
Remark
The
AC
setting sets the Aux Trig Out to be the probe compensation square wave
(approximately 750 Hz).
CAL
Space
?
:OUTP
character_data
Item
Description/Default
Range of values
character_data
Sets the calibration output as:
–
AC
: Sets the AC coupling frequency.
–
DC,<dc_value
>: Sets the DC level, with its value in volts
within the range of –2.4 V to +2.4 V.
AC
DC,<dc_value>
CAL:OUTP DC,2.0
This command puts a DC voltage of 2.0 V on the Aux
Trig Out connector..