XYDISPlay Subsystem
17
Keysight 8990B Programming Guide
465
XYDISPlay:XSOURce <character_data>
This command sets the X-axis source for the XY Display mode. The X-axis scale
depends on the X-axis source.
Syntax
Parameter
Example
Remark
This command is only applicable in the
XYDISPlay
acquisition mode. If the
acquisition mode is not set to
XYDISPlay
, error –221,“Settings conflict;Requires
XY display mode to be enabled” will occur. Refer to
on page 132 to set the acquisition mode.
Reset condition
On reset, the X-axis source for the XY Display mode is set to CHAN1.
:XSOUR
Space
character_data
XYDISP
Item
Description/Default
Range of values
character_data
Sets the XY Display X-axis source as:
– CHANnel<n>
– WMEMory<n>
– FUNCtion{1|2|3|4}
where n = 1, 2, 3, or 7
CHANnel<n>
WMEMory<n>
FUNCtion
{1|2|3|4}
XYDISP:XSOUR
CHAN2
This command sets the X-axis source for the XY Display
mode to channel 2.