TRIGger Subsystem
8
Keysight 8990B Programming Guide
207
TRIGger:HYSTeresis:LEVel <numeric_value>
This command sets the trigger hysteresis level for channels 1 and 4.
Syntax
Parameter
Example
Remarks
– This command is only applicable for channels 1 and 4. If you send this
command for channels other than 1 and 4, error 705,“Applicable to trigger
source for channel 1 and 4 only” will occur.
– If you set a hysteresis level value which exceeds its minimum or maximum
limit, the value will be clipped to its minimum or maximum value respectively.
Error –222,“Data out of range;Value clipped to minimum (#)” or –222,“Data out
of range;Value clipped to maximum (#)” will occur.
TRIG
Space
?
numeric_value
:HYST
:LEV
Item
Description/Default
Range of values
numeric_value
A numeric value in dBm for the hysteresis level:
– Default value: 0 dBm
– Minimum value: 0 dBm
– Maximum value: 3 dBm
0 to 3 dBm
TRIG:HYST:LEV 1
This command sets the trigger hysteresis level to
1 dBm.