MEASure Subsystem
9
Keysight 8990B Programming Guide
263
MEASure:THReshold:PDURation CHAN{1|2|3|4},<numeric_value>
This command sets the pulse duration level for any of the four channels.
Syntax
Parameter
Example
NOTE
– The minimum value for the pulse duration is clipped to the value of reference
level 1 incremented by one when the marker source is set to one of the four
channels.
– The maximum value for the pulse duration is clipped to the value of reference
level 2 decremented by one when the marker source is set to one of the four
channels.
Item
Description/Default
Range of values
numeric_value
A numeric value in % for the pulse duration level:
– Default value: 50%
– Minimum value: 0%
– Maximum value: 100%
0 to 100%
:THR
?
MEAS
Space
CHAN
1
2
3
4
,
numeric_value
:PDUR
MEAS:THR:PDUR CHAN1,60
This command sets the pulse duration level
to 60% for channel 1.