4
ACQuire Subsystem
116
Keysight 8990B Programming Guide
Remarks
– This command is only applicable for channels 1 and 4. If you send this
command for a channel other than 1 or 4, error 700,“Applicable to channel 1
and 4 only” will occur.
– This command is only applicable in the CCDF acquisition mode. If the
acquisition mode is not set to CCDF, error –221,“Settings conflict;Requires
CCDF mode to be enabled” will occur. Refer to
on page 132 to set the acquisition mode.
– If you set a scale value which exceeds its minimum or maximum limit, the value
will be clipped to its minimum or maximum value respectively. Error
–222,“Data out of range;Value clipped to minimum (#)” or –222,“Data out of
range;Value clipped to maximum (#)” will occur.
Reset condition
On reset, the CCDF trace display scale is set to 5 dB/div.
Query
ACQuire:CCDF:SCALe?
This query returns the current setting of the CCDF trace display scale. The
response format is
<NRf>
.
Query example
ACQ:CCDF:SCAL?
Queries the CCDF trace display scale setting.