SYSTem Subsystem
22
Keysight 8990B Programming Guide
537
–
222
Data out of range
A numeric parameter value is outside the valid range for the command
Example: CHAN1:EXT 500
–
222
Data out of range;Value clipped to minimum (#)
A numeric parameter value is lower than the minimum value allowed
–
222
Data out of range;Value clipped to maximum (#)
A numeric parameter value is higher than the maximum value allowed
–
223
Too much data
A data element was received that contains more data than the 8990B can handle
–
224
Illegal parameter value
A discrete parameter was received which was not a valid choice for the command. You may have used an
invalid parameter choice.
Example: CHAN1:REF MIDD
–
225
Out of memory
The 8990B has insufficient memory to perform the requested operation
–
231
Data questionable;Calibration error
8990B calibration and zeroing has failed
–
231
Data questionable;Voltage overloaded
The voltage input to the channel exceeds the maximum range of the probe
–
231
Data questionable;Calibration error in channel 1
An error has occured during calibration for channel 1
–
231
Data questionable;Calibration error in channel 4
An error has occured during calibration for channel 4
–
232
Invalid format
The data format or structure is inappropriate
–
233
Invalid version
The version of the data format is incorrect
–
240
Hardware error
The command could not be executed due to a hardware problem
–
241
Hardware missing;Unable to turn on channel #;no sensor detected
The 8990B is unable to execute the command because no sensor is connected to the respective channel
Summary of Contents for 8990B
Page 1: ...Keysight 8990B Peak Power Analyzer Programming Guide ...
Page 16: ...16 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...20 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...