6
CHANnel Subsystem
164
Keysight 8990B Programming Guide
Remarks
– This command is only applicable for channels 1 and 4. If a channel other than
channel 1 or 4 is specified, error 700,“Applicable to channel 1 and 4 only” will
occur.
– If you set an external loss value which exceeds its minimum or maximum limit,
the value will be clipped to its minimum or maximum value respectively. Error
–222,“Data out of range;Value clipped to minimum (#)” or –222,“Data out of
range;Value clipped to maximum (#)” will occur.
– If channel 1 or 4 is not connected, error –241,“Hardware missing; Sensor is not
found in channel <channel>.” will occur.
Reset condition
On reset, the external loss is set to 0 dB.
Query
CHANnel{1|4}:EXTLoss?
This query returns the current setting of the external loss for the specified
channel. The response format is
<NRf>
.
Query example
CHAN4:EXTL?
Queries the external loss setting for channel 4.