6
CHANnel Subsystem
168
Keysight 8990B Programming Guide
Remarks
– If an invalid table name is specified, error –224,“Illegal parameter value” will
occur.
– This command is only applicable for channels 1 and 4. If a channel other than
channel 1 or 4 is specified, error 700,“Applicable to channel 1 and 4 only” will
occur.
– If there is no data in the selected FDO table, error –221,“Settings conflict;No
FDO data entry available” will occur.
Query
CHANnel{1|4}:FDOffset:SELect?
This query returns the current setting of the FDO table for the specified channel as
a string value. An empty string "" will be returned if no table has been selected.
Query example
CHAN1:FDO:SEL?
Queries the FDO table assigned to channel 1.