ACQuire Subsystem
4
Keysight 8990B Programming Guide
135
ACQuire:RF:INTerpolate:RESolution <character_data>
This command sets the resolution during interpolation in number of points per
nanosecond.
Syntax
Parameter
Example
Reset condition
On reset, the interpolation algorithm resolution is set to
POINT1N
.
NOTE
This command can only be used when
ACQuire:RF:INTerpolate
is set to
AUTO
(refer to
ACQ
:RF
Space
:INT
character_data
:RES
Item
Description/Default
Range of values
character_data
Sets the interpolation algorithm resolution to:
–
POINT1N
: Sets the resolution to 1 point per nanosecond.
–
POINT2N
: Sets the resolution to 2 points per nanosecond.
–
POINT5N
: Sets the resolution to 5 points per nanosecond.
POINT1N
POINT2N
POINT5N
ACQ:RF:INT:RES POINT2N
Sets the interpolation algorithm resolution to 2
points per nanosecond.