TRIGger Subsystem
8
Keysight 8990B Programming Guide
205
TRIGger:HOLDoff <numeric_value>
This command sets the trigger holdoff value.
Syntax
Parameter
Example
Remark
If you set a trigger holdoff value which exceeds its minimum or maximum limit, the
value will be clipped to its minimum or maximum value respectively. Error
–222,“Data out of range;Value clipped to minimum (#)” or –222,“Data out of
range;Value clipped to maximum (#)” will occur.
TRIG
Space
?
numeric_value
:HOLD
Item
Description/Default
Range of values
numeric_value
A numeric value for the trigger holdoff:
– Default value: 1
μ
s
– Minimum value: 1
μ
s
– Maximum value: 1 s
1
μ
s to 1 s
TRIG:HOLD 0.5
This command sets the trigger holdoff to 0.5 s.