11
MARKer Subsystem
324
Keysight 8990B Programming Guide
Parameter
Example
Remarks
– This command is only applicable for the marker spacing measurement. If this
command is sent for a marker measurement mode other than spacing, error
–221,“Settings conflict; Requires to switch to pulse spacing marker
measurement” will occur.
– If you set a pulse spacing value which exceeds its minimum or maximum limit,
the value will be clipped to its minimum or maximum value respectively. Error
–222,“Data out of range; Value clipped to minimum (#)” or –222,“Data out of
range; Value clipped to maximum (#)” will occur.
Item
Description/Default
Range of values
numeric_value
A numeric value for the lower and upper pulse spacing:
– Default values:
– Lower pulse spacing (
PULSe1
): 1 (represents the 1st pulse)
– Upper pulse spacing (
PULSe2
): 1 (represents the 1st pulse)
– Minimum value: 1 (represents the 1st pulse)
– Maximum value: the maximum edge captured in an acquisition
1 to (the maximum edge
captured in an acquisition)
MARK:SPAC:PULS1 10
MARK:SPAC:PULS2 20
These commands set the marker pulse spacing
for spacing measurement in between the 10th
pulse and the 20th pulse.