ACQuire Subsystem
4
Keysight 8990B Programming Guide
83
ACQuire:AVERage:COUNt <numeric_value>
This command sets the filter length for the averaging operation.
Syntax
Parameter
Example
Remarks
– If the averaging state is off, it will be turned on when you set the averaging
filter length.
– If you set a value other than the range specified for the filter length, error
–222,“Data out of range” will occur.
ACQ
:AVER
Space
?
:COUN
numeric_value
Item
Description/Default
Range of values
numeric_value
A numeric value for the averaging filter length:
– Default value: 2
– Minimum value: 2
– Maximum value: 2048
2, 4, 8, 16, 32, 64, 128, 256,
512, 1024, or 2048
ACQ:AVER:COUN 128
This command sets the averaging filter length to