ACQuire Subsystem
4
Keysight 8990B Programming Guide
111
ACQuire:CCDF:REFerence[:STATe] <boolean>
This command sets the state of the CCDF reference trace.
Syntax
Example
Remarks
– This command is only applicable for channels 1 and 4. If you send this
command for a channel other than 1 or 4, error 700,“Applicable to channel 1
and 4 only” will occur.
– This command is only applicable in the CCDF acquisition mode. If the
acquisition mode is not set to CCDF, error –221,“Settings conflict;Requires
CCDF mode to be enabled” will occur. Refer to
on page 132 to set the acquisition mode.
– This command is only applicable when channel 1 or 4 is connected. If the
channel trace is not connected, error –221,“Settings conflict;Requires CCDF
trace # to be enabled” will occur. Refer to
“ACQuire:CCDF:TRACe{1|4}[:STATe]
on page 125 to set the CCDF channel trace.
ACQ
:STAT
:CCDF
0|OFF
1|ON
Space
?
:REF
ACQ:CCDF:REF 1
This command turns on the CCDF reference trace.
Summary of Contents for 8990B
Page 1: ...Keysight 8990B Peak Power Analyzer Programming Guide ...
Page 16: ...16 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...20 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...