ETS <character_data>
This command sets the state of the Equivalent-Time Sampling (ETS) event.
Syntax
Parameter
Example
Remarks
– Turning on ETS enables the 8990B to capture high-frequency signals with a
higher effective sampling rate.
– Video bandwidth is turned off when ETS is enabled.
– The 10 MHz reference signal input is turned off when ETS is enabled. If ETS is
enabled while the 10 MHz input is turned on, error 939, “10 MHz setting is only
applicable when ETS is off. 10 MHz setting is turned off” will occur.
– During AUX trigger mode, pressing the single button will ensure that ETS
acquisition is completed before returning the trace. As compared to other
trigger sources, the user needs to send multiple single commands or press the
single button to get the completed ETS trace.
Space
?
ETS
character_data
Item
Description/Default
Range of values
character_data
Sets the ETS state as:
–
AUTO
: Sets ETS to automatically turn on when the timescale
value is less than the value set at ETSThreshold.
–
OFF
: Turns off ETS.
AUTO
OFF
ETS OFF
This command turns off the ETS event.