FluoroMax-3 v. 3.1 (3 Mar 2006)
Troubleshooting
7-5
Note:
Not all spectra shown in this section were produced using the
FluoroMax
®
-3. The spectra are presented to show different possible
system or sample problems, and may not reflect the performance of
your particular FluoroMax
®
-3.
Using diagnostic spectra
Often the spectrum reveals information regarding the hardware or software parameters
that should be adjusted. The following spectra occur with explanations about problems
leading to their appearance.
Lamp scan
Running a lamp scan verifies system integrity and indicates whether the correct pa-
rameters for the best possible trace are being used. The following spectrum shows the
trace resulting from a lamp scan run with a known good lamp.
0
0.01
0.02
0.03
0.04
0.05
0.06
0.07
250
300
350
400
450
500
550
600
In
te
n
s
it
y
W avelength (nm )
467 nm
Scan of good quality 150-W xenon lamp in FluoroMax
®
-3.