FluoroMax-3 v. 3.1 (7 Mar 2006)
Components & Accessories
12-13
Note:
Selecting too high a
speed may create a vortex,
which could affect the re-
producibility of the meas-
urement.
Note:
While the four-position
model maintains the tempera-
ture of all four samples, only
one sample is mixed at a time.
Use
1
Place the sample in a 10 mm × 10 mm cuvette
and insert a magnetic stirring bar.
(The stirring bar is available from Bel-Art Products, Pequannock, NJ)
2
Place a cuvette in
each holder.
3
Allow the
samples to reach
the desired temperature.
4
Turn on the magnetic stirrer.
5
Select the
appropriate mixing
speed.
The speed at which the sample
should be mixed depends on the
viscosity of the sample.
6
Run your experiment as usual.
7
Place the next cuvette in the sample position by
lifting up the knob and rotating the holder.
Be sure to press down, to lock the cuvette into the proper position.