FluoroMax-3 v. 3.1 (3 Mar 2006)
Optimizing Data
5-7
Note:
For weak signals, enter the
G
factor, rather than measure it auto-
matically. This may improve the S/N.
Note:
For detailed information on the
G
factor, see the
Polarizers
Operation Manual
.
1
In the Experiment Setup window, click the De-
tectors icon.
This shows the parameters related to detectors, including the
G
factor, in the
Polarization
area.
2
Click the G Factor checkbox to include a
G
fac-
tor in your measurements.
3
Enter a value for the
G
factor in the field if you
want to use a pre-determined value. To auto-
matically measure the
G
factor, use the default
value of 0.